Breathing hose
•
We recommend GIGASEPT FF as a disinfectant.
•
When disinfecting use the same procedure as for cleaning.
•
The breathing hose must not be heated to temperatures in excess of
70 °C. Sterilization is not permitted.
Accessories
To disinfect or sterilize the exhalation systems and the mask, please see the section "Hy-
gienic preparation" in the relevant operating instructions.
When there is a change of patient please refer to section "5.6 Disinfection CLICK 2" on
page 30 and "5.8 Multiple patient use CLICK 2" on page 31 for instructions on how to dis-
infect the CLICK 2 humidifier.
5.6 Disinfection CLICK 2
The humidifier can be disinfected. Please observe the operating instructions for the disin-
fectant used. We recommend that you wear suitable gloves when disinfecting the equip-
ment (e.g. household or disposable gloves).
We recommend GIGASEPT FF.
•
The procedure for dismantling and assembling the humidifier is the same
as for cleaning.
•
To ensure full disinfection, the filling cap should also be removed and dis-
infected separately.
For hygiene reasons, we recommend that the plastic parts are replaced after a maximum
of two years use.
5.7 Multiple patient use CPAP 20e
If the unit is to be utilized for another patient without using a bacteria filter, it must under-
go a hygiene treatment beforehand. This has to be carried out by Weinmann or a dealer.
The procedure used for the hygiene treatment is described in the inspection sheet and the
service and repair instructions for CPAP 20e.
30
Hygienic preparation