3.3.2
Switching on the unit
After connecting the unit to the power supply turn on the power by
setting the power switch (5) to the "I" position (the switch will light
up).
When the power is turned on, the unit will indicate the last
sterilization mode performed and the alarm "Please open the door
before working" will sound if the door is closed before the power is
turned
Open the door and you can operate the unit further, i.e. go to the
desired
3.3.3
Changing the settings
Display:
•
Clock
•
Date
Pressure in the chamber (PRESSURE)
•
•
Temperature in the chamber (TEMPERATURE)
•
Current operating mode (MODE)
•
Current status (STATE)
Total number of running cycles (CYCLE)
•
Heating ring / evaporator temperature (OUT-T /
•
STE-T)
Distilled water tank and main tank level (WATER1 /
•
WATER2)
Door lock status (DOOR)
•
Error code (ERROR)
•
Buttons:
a) Menu / parameter setting button
b) Button to scroll up an item/parameter
c) Button to scroll down the item/parameter
d) Button for accepting the parameter and starting or
stopping the unit
on.
mode.