Tailoring treatment setup options
Setting Ramp
Programs
Note: Each program retains its own circuit configuration. When switching between programs, ensure you
use the correct circuit (breathing system) as 'learned' for that program. If more than one program has been
set up, follow the directions given by your clinician for when and how each program should be used.
Ramp Down
Using mask-fit
You can use mask-fit to help you fit your mask properly. This feature delivers constant treatment pressure
for a three-minute period, prior to starting treatment, during which you can check and adjust your mask-fit
to minimise leaks. The mask-fit pressure is the set CPAP or EPAP pressure or 10 cm H
greater.
1. Put the mask on according to the mask user guide.
2. Hold down
for at least three seconds until pressure delivery starts.
Ramp time is a feature that can be enabled by your clinician by
setting a maximum ramp time. Designed to make the
beginning of treatment more comfortable, ramp time is the
period during which the pressure increases from a low start
pressure to the treatment pressure. Refer to Setup menu:
Options (see page 19).
Programs can be configured by your clinician to provide you with
multiple treatment options. For example, a clinician can set up
programs for sleeping versus waking use or for use during
exercise or physiotherapy. Programs save different patient
circuit configurations (Learn Circuit result) and therapy and alarm
settings.
The Stellar comes with one active program. Your clinician can
configure two programs. If your clinician has selected dual, you
can choose the program to use on the Treatment screen after
stopping treatment. If only a single program is selected, the
option does not display.
If you experience difficulties in taking off the ventilation, you can
use the Ramp Down feature. This feature reduces the treatment
pressure to Start EPAP/PEEP over five minutes. After five
minutes the therapy stops. You can enable the Ramp Down
feature when it is selected by the physician in the Options
menu.
Notes:
•
Therapy can be stopped at any time by pressing the therapy
start/stop button, including during the Ramp Down time.
•
During the Ramp Down time the low pressure alarm is not
active.
•
This feature is not available in all countries.