Processing
Instructions apply to the sterilization tray only. For instructions on how to
process the devices, consult their respective instructions for use.
Point of Use
Containment and
Transportation
Preparation for
Cleaning
Pre-Treatment
Caution
• Wipe excess soil from the tray.
• Prevent soil from drying by keeping device
moist until ready for further processing.
• Follow the facility's internal procedures for
the transportation of contaminated surgical
instruments and devices.
• Process device(s) / tray as soon as possible
following use.
• Disassemble the tray into its individual
components: base, lid, and (if applicable)
silicon mat.
For necessary materials and equipment, see
Materials and Equipment table.
1.
Rinse device under cool, running utility water
to remove gross soil from device. Ensure all
external and any accessible internal surfaces
are rinsed.
2.
Drain excess water from the device.
3.
Follow instructions for manual or automated
cleaning.
• Do not use brushes or pads with metal or
abrasive materials to clean or disinfect the
devices; as permanent scoring or damage
could result.
• Do not allow devices to remain in any solution
longer than the manufacturer's indicated times;
as corrosion or damage could result.
• Exposure to excessive pH will strip protective
coatings from metals, especially aluminum.
EN-7