Data management
There are two data ports at the rear of the device for connecting a USB stick. Refer to Stellar at a glance
(see page 2). Therapy and device data can be stored on it or read from it to be used with ResMed software
applications.
WARNING
Do not connect any device to the data ports other than specially designed devices recommended by
ResMed. Connection of other devices could result in injury, or damage to the Stellar. Refer to
General warnings and cautions (see page 32).
CAUTION
Do not disconnect the ResMed USB stick while data transfer is in progress. Otherwise loss of data or
incorrect data may result. The download time depends on the data volume.
Notes:
•
You cannot connect two ResMed USB sticks simultaneously for data communication.
•
If data transfer is not possible or failed please read the troubleshooting section.
•
Do not store files on the USB stick other than those created by the device or application. Unknown files
may be lost during data transfer.
1. Plug the ResMed USB stick to one of the two USB connections on the rear side of the device.
The first dialogue for the USB data transfer is displayed automatically on the LCD. The device checks if
there is enough memory capacity on the USB stick and if readable data is available.
2. Select one of the available options
• Read settings
The settings will be transfered from the connected ResMed USB stick to the Stellar.
• Write settings
Settings of the device will be stored on the ResMed USB stick.
• Write settings and loggings
Settings and loggings of the device will be stored on the ResMed USB stick.
• Cancel
3. Confirm the data transfer.
Cleaning and maintenance
The cleaning and maintenance described in this section should be carried out regularly. This also helps to
prevent the risk of cross contamination. Refer to the mask, humidifier and other accessories user guide for
detailed instructions for care and maintenance.
WARNING
•
Beware of electric shock. Do not immerse the device, pulse oximeter or power cord in water.
Turn off the device, unplug the power cord from the power socket and the device before cleaning
and be sure that it is dry before reconnecting.
•
The mask system and air tubing are subject to normal wear and tear. Inspect them regularly for
damage.
CAUTION
The Stellar cannot be sterilised.
Daily
Disconnect the air tubing from the device (and humidifier, if used) and hang it in a clean, dry place until
next use. If the device is visibly soiled, wipe the exterior surfaces of the device and the pulse oximeter (if
used) with a damp cloth and mild detergent.