Table 2 – Guidance And MANUFACTURER'S Declaration
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The EGA-500/EGA-500E is intended for use in the electromagnetic environment specified below. The
customer or user of the EGA-500/EGA-500E should ensure that it is used in such an environment.
Immunity Test
ESD
EN/IEC 61000-4-2
EFT
EN/IEC 61000-4-4
±1kV Differential
Surge
EN/IEC 61000-4-5
Voltage Dips/
Dropout
EN/IEC 61000-4-11
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-4-8
IFU-51 / REV G
Electromagnetic IMMUNITY
For All ME EQUIPMENT And ME SYSTEMS
EN/IEC 60601
Compliance
Test Level
Level
±6kV Contact
±6kV Contact
±8kV Air
±8kV Air
±2kV Mains
±2kV Mains
±1kV I/Os
±1kV I/Os
±1kV Differential
±2kV Common
±2kV Common
>95% Dip for
>95% Dip for
0.5 Cycle
0.5 Cycle
60% Dip for
60% Dip for
5 Cycles
5 Cycles
30% Dip for
30% Dip for
25 Cycles
25 Cycles
>95% Dip for
>95% Dip for
5 Seconds
5 Seconds
3A/m
3A/m
Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic tile. If floors
are synthetic, the r/h should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment. If the user
of the EGA-500/EGA-500E requires continued operation
during power mains interruptions, it is recommended
that the EGA-500E be powered from an uninterruptible
power supply or battery.
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.
EN - 16