The torque setting can be adjusted from 0.6 to 4.0 Ncm in increments of 0.1 Ncm.The preset
torque and speed values of all continuous rotary file systems can be altered individually.
CAUTION
Before using the motor handpiece, verify the correctness of the changed parameters.
6.8.

Factory Default Parameters

To return to the original default parameters, follow the general reset instructions:
1. Turn the power off.
2. Connect the AC adapter and check that the AC-IN lamp lights (see chapter 6.5.1
Connecting AC adapter).
3. Hold down the POWER key for more than 2 seconds, while pressing the MEMO key.
• During the process the display will read:
• When the process is completed, the display reads:
• Then, the display returns to the first system of the file library.
NOTES
• This function is not activated unless the device is powered by the AC adapter.
• Be aware that all individual settings will be deleted when "reset to default parameters" has
been carried out.
FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
- 28 / 208 -