DENTSPLY x-smart plus Manual Del Usuario página 36

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Error
code
During
E-00
rotation of
E-01
the motor
handpiece
E-02
E-03
E-04
E-05
E-06
E-08
E-09
At the time
E-10
of charging
E-11
E-12
E-13
E-14
FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
Error
Self-Check
Malfunction of circuit.
Over current
The motor handpiece is locked
(at the time of the auto reverse
mode).
The motor handpiece cord has
shorted.
Over voltage
Malfunction of circuit.
Motor handpiece
The motor handpiece cord is
sensor
disconnected.
Faulty sensor (Hall IC).
Severed cord (signal line).
Overheating of
High load was continuously
motor
applied to the motor handpiece
for a relatively long period of
time.
PAM circuit
Abnormal voltage generated in
start / stop circuit.
Faulty start / stop circuit from
PAM (L Slide).
Rotor lock
The motor handpiece is locked
at the time of startup.
Faulty contra-angle.
Motor faulty.
Faulty sensor (Hall IC).
Severed cord (Signal, Power
line).
Over current
Shorted cord (power line).
Shortstop of the motor winding.
ITRIP
Faulty motor and circuit.
Battery current
The battery current is too low
or too high. The battery is flat
or not inserted.
Display
Faulty display driver.
Low battery voltage
The battery voltage is too low.
The battery is not inserted or
the battery operating life is
over.
High battery
The battery voltage is too high
voltage
(circuit malfunction).
Beyond the range
Beyond the range of working
of working
temperature or damage to the
temperature
thermistor in the battery
section.
Cause
Contact your distributor.
Remove load.
Contact your distributor.
Contact your distributor.
Connect the motor
handpiece cord correctly.
Contact your distributor.
Let the motor cool down
before resuming
operation.
Contact your distributor.
Remove load.
Contact your distributor.
Contact your distributor.
Contact your distributor.
Put the battery into the
battery compartment or
replace the battery.
Contact your distributor.
Put the battery into the
battery compartment or
replace the battery.
Contact your distributor.
Use within the range of
working temperature or
replace the battery.
Check/Remedy
- 37 / 208 -