9.3 Static Alignment
1 Complete the Bench Alignment set-up (see Section 7 )
2 Click on
A
to select the Dynamic Alignment Vector screen.
A
3 Static Alignment Procedure
With the user standing comfortably adjust
the Elan A-P shift to achieve a neutral
balance position*.
See following pages.
Tilt
B
Use the display to adjust the angular
position of the Elan when standing.
Adjust using shift and/or tilt devices as
required.
( When standing the Vector indicator
position shown is recommended)
Warning:
Ensure the user is not resting on the DF
(Dorsiflexion) limit when standing in an
upright position.
* Ensure the Elan is uncalibrated prior to
static alignment. See Section 9.5 Factory
Reset.
Alignment
B
Tilt setting position
15
3°
6°
938317S/3-0719