Status Bar
The current status of the Elan is shown by the indicators on the Status bar.
Mode Indicator
Mode
Angle
Tilt
PF and DF Valve
Battery
Power Save/ Sleep
Standing Support*
Bluetooth
Power Save/ Sleep
After Calibration if the Elan detects no movement for approximately 10 seconds it
activates power saving mode disabling the sensors. Note during the calibration and
set-up the time interval is increased to 90 seconds.
*Standing Support Mode
This mode is only available if enabled, the default setting is OFF see Section 14 Options.
When the Elan detects a stationary standing posture high ankle resistance is selected.
To adjust or fine tune the setting on a
previously calibrated Elan:
If the alignment has not been changed it is not necessary to follow the whole set up
procedure again. Follow the Walking Calibration sequence (see section 9.5) and the
Elan software will then add the new information to the previous data and average out
the result giving improved settings for the user.
If alignment adjustments are required carry out a full factory reset and follow the full
set up procedure.
Description
Displays the current Elan mode
Shows foot angle dynamically
Shows tilt position dynamically
Displays the current dynamic resistance settings for plantar
flexion and dorsiflexion
shows the Elan is powered up
If displayed the Elan is in Power Save mode*
z
z
z
If displayed Standing Support Mode is active,
high ankle resistance is selected.
Shows that the Elan is connected/disconnected via Bluetooth
14