Walking Calibration procedure (continued)
2 Calibration stage 2 (assist)
Instruct the user to walk on a horizontal
and level surface and at a comfortable
pace and after several steps,
click on:
Calibration
E
The steps counter shows the number of
steps remaining.
3 Calibration stage 3 (brake)
Instruct the user to walk on a horizontal
and level surface and at a comfortable
pace and after several steps,
click on:
Calibration
F
This completes the Elan calibration.
Click on:
[Logout] and [X] to exit the Program.
D
stage 2 (assist
D
stage 3 (brake)
E
F
Stages 2 and 3 will feel unfamiliar;
this is normal and will pass when
completed.
)
Factory Reset:
Should it be necessary to begin a
completely new set-up for the Elan,
Clicking on [Factory Reset] will
delete all the Dynamic Alignment
adjustment and any walking
calibration data stored in the Elan.
H
Click on
If this option is chosen return to
Section 9.4 to begin set-up.
31
H
Caution:
Factory Reset
938317S/3-0719