en
Disinfection
Genius 3 Thermometer Body, Base, Coil Cord, Probe Tip and Lens:
The Genius 3 thermometer body, base, coil cord, probe tip, and lens can be disinfected by dampening their surfaces with 70%
isopropyl alcohol.
Disinfection frequency:
It is necessary to disinfect the Genius 3 thermometer after each use.
Directions for Disinfecting the Thermometer Body, Base, Coil Cord, Probe Tip, and Lens:
Dampen the surfaces of the devices using isopropyl wipes such as Webcol™*, Curity™*, or an equivalent lint free wipe wetted with
70% isopropyl alcohol. Wipe the devices as necessary to maintain visual wetness for a minimum duration of 1 minute. Ensure that
the thermometer lens is free from fingerprints and / or smudges for proper operation. After disinfecting the thermometer probe
tip and lens, allow the thermometer to air dry completely.
Caution: The Genius 3 Tympanic Thermometer and Base is a non-sterile device. Do not use ethylene oxide gas, heat, autoclave, or
any other harsh method to sterilize this thermometer.
Caution: The Genius 3 Tympanic Thermometer and Base are not designed to be immersed, soaked, rinsed, or sprayed with water.
Do not immerse, soak, rinse, or spray the Genius 3 Tympanic Thermometer or Base in water or other cleaning solutions. Failure
to follow the cleaning procedures described herein could result in hazards to users, patients, and clinicians. As with any medical
electrical equipment, care must be taken to prevent liquid from entering the thermometer to avoid electrical shock hazard, fire
hazard, or damage to the electrical components.
Caution: If leakage into the thermometer interior occurs, do not use the thermometer until it has been properly cleaned,
dried, and checked for accuracy. Accuracy can be verified with the Genius Checker/Calibrator. For assistance please contact
Customer Service.
Section VII — Battery Replacement
The batteries (3 AAA) should be changed when the low battery icon appears on the display. After the low battery icon is shown,
about 100 readings can be taken before the dead battery display will appear. Once the dead battery display is shown, no more
readings can be taken.
To change the batteries, remove the cover on the bottom of the unit. Note the polarity of the installed batteries. Remove the old
batteries and replace with new, using the correct polarity. Replace the battery cover and secure it with the screws.
Section VIII — Mounting Instructions
A wall mount or cart mount (sold separately) are available for the base unit. Follow the instructions supplied with those items.
Tympanic Thermometer and Base
10
Table of Contents