Biotech Mode
Biotech Mode contains the site mode options and can display the installed software version. All site mode settings in Biotech
Mode are retained through power cycles, such as changing batteries.
The factory default settings are shown below:
Temperature mode °C (unlocked)
Site mode
Ear
Site text
On
To enter Biotech Mode the thermometer must first be in "off" mode or sleep mode. While the thermometer is in "off" mode
or sleep mode, press and hold the timer and °C/°F buttons for four seconds. All LCD segments will light for one second, the
thermometer will beep once, and the display will show scrolling dashes. Press the timer button to cycle through the biotech
modes. When options are available within a mode, the °C/°F button cycles through the options.
Pressing the timer button after the site text display will return the user to the installed software version.
The device will exit Biotech Mode after 30 seconds of non-use. To manually exit Biotech Mode press and hold the °C/°F and timer
buttons for one second. Any changes are saved.
The biotech mode sequence is shown below:
Software version
Displays the installed software version of device. Where "00" is the current software version.
Temperature mode
°C (unlocked)
Site mode
Ear
Tympanic Thermometer and Base
Locked °C
Oral
Locked °F
Rectal
8
°F (unlocked)