Section V Preventative Maintenance; Section Vi Cleaning & Disinfecting - Covidien Genius 3 Manual De Instrucciones

Ocultar thumbs Ver también para Genius 3:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 85
en
Site text
Pressing the °C/°F button when in this mode turns the body site text labels on or off. The labels will remain on when an "X" appears
inside the box icon, and the text will remain off when the box is empty.
Section V — Preventative Maintenance
A checker/calibrator is available for this device or the device can be sent in for service. The device must be checked for calibration
every 25 weeks or whenever calibration is in question. If not able to calibrate, contact your Covidien representative for details.
Harsh use or harsh environmental conditions may result in the need for more frequent checks. If the unit is dropped, abused, or
stored at less than -25°C, or above 55°C, check it before the next use.
Section VI — Cleaning & Disinfecting
Cleaning
Genius 3 Thermometer Body, Base, and Coil Cord:
A mild, common dish washing liquid detergent, such as Dawn®, should be used for general cleaning of the thermometer body,
base, and coil cord. This detergent should be used with a 20:1 ratio water to detergent mixture. The water and detergent mixture
should not exceed 55°C (130°F).
Caution: The probe tip of the Genius 3 thermometer should not be cleaned with mild detergent.
Genius 3 Thermometer Probe Tip and Lens:
The probe tip and lens can be cleaned using a 70% Isopropyl alcohol wipe such as Webcol™*, Curity™*, or equivalent.
Caution: The use of other cleaners and disinfectants may cause significant damage to the Genius 3 thermometer and base
and may void warranty. Never use an abrasive pad on any surface of the Genius 3 thermometer.
Cleaning frequency:
It is recommended that the Genius 3 thermometer, base, and coil cord be cleaned after each use.
Directions for Cleaning:
Genius 3 Thermometer Body, Base, and Coil Cord:
While cleaning the Genius 3 thermometer body, a probe cover should be installed on the thermometer. This will prevent
damage to the probe tip and probe lens. Clean the surfaces of the thermometer body, base, and coil cord with a damp cloth
using a mild detergent as previously described, removing all visible soil. Ensure all excess fluid is squeezed from the cloth before
cleaning. If the cloth is excessively wet, the detergent and water solution may penetrate the thermometer and affect functionality.
After cleaning the thermometer body, base, and coil cord wipe the thermometer with a clean lint free damp cloth to remove the
mild detergent mixture. Dry the thermometer with a clean lint free cloth. Never use an abrasive pad or an abrasive cleaner on the
Genius 3 thermometer, base, or coil cord.
Genius 3 Thermometer Probe Tip and Lens:
The thermometer probe tip and lens can be cleaned using a 70% isopropyl wipe such as Webcol™*, Curity™* or equivalent.
Carefully remove all foreign matter from the thermometer probe tip and lens. After the foreign matter has been removed, dry
the lens at the end of the thermometer probe tip with a lint free swab, cotton ball, or lens wipe. The thermometer lens must
be free from fingerprints and / or smudges for proper operation. After cleaning the thermometer probe tip and lens, allow the
thermometer to air dry completely.
Tympanic Thermometer and Base
9
Table of Contents
Tabla de contenido
loading

Tabla de contenido