Arthrex Shoulder Suspension System AR-1650S
Instructions for Use
Mount Clamp to the rail by sliding it on the distal end of the rail.
Typically, the Clamp is placed on the contralateral side of the patient
but it is not required. It is recommended for most tables and patients
that the clamp be pulled back to engage the rail gravity lock. Secure the
Table Clamp to the rails by tightening the two wing knobs on the Clamp
body
(Figure
1).
Lift the lever, move it approximately to center and push the lever down
(Figure
2). Adjusting the lever forward or backward will adjust the angle
of the Tower relative to the OR table. It is recommended that the
starting orientation is perpendicular to the rail in order to mount the
Tower parallel with the patient but it is not required.
Manufactured For:
Arthrex, Inc.
1370 Creekside Blvd
Naples, FL 34108, USA
1-(800) 934-4404
Figure 1
Figure 2
Arthrex GmbH
Allen Medical Systems
Erwin-Hielscher-Strasse 9
100 Discovery Way
Acton, MA 01720
81249 München, Germany
(978) 263-5401
+49 89 909005-0
Suspension Tower
Remove the Suspension Tower from the Storage Cart or designated
storage area by holding it by the blue carrying handles shown in
Place the hook end of the Tower (AR-1650-01) into the slot in the stem
of the clamp as shown
(Figure
pivot it in the Clamp
(Figure
4). The Tower will rest at about 15 degrees
toward the head of the patient.
Hill Rom SAS
B.P. 14 Z.I. du Talhouet
56330 Pluvigner, France
Tel. +33 (0) 2 97 50 92 12
Fax. +33 (0) 2 97 50 92 12
Figure 3
3). Lift and push forward on the Tower to
D-720764-A2
3 AUG 2016 Page | 3
Figure
3.