2
Installation
This chapter covers
•
Opening the Enclosure
•
Environmental Protection
•
Mounting the Terminal
•
Installing Cables and
Connections
•
Capacity Label Instructions
•
Closing the Enclosure
•
Sealing the Enclosure
2.1.
Opening the Enclosure
The front panel of the IND231/IND236 terminal is locked in place by four screws which attach it to
the rear housing of the enclosure. To access the terminal's PCB in order to install options, connect
internal wiring and set switches, separate the front panel from the enclosure as follows:
1. For IND231, use a T-20 torx screwdriver to loosen three of the screws. Use a slotted
2. For IND236, use a 10mm socket wrench to loosen three of the screws. Use a 5mm Allen Head
30094803
|
00
|
08/2013
This chapter provides installation instructions for the IND231/IND236 terminal.
Please read these procedures thoroughly before beginning installation.
This section includes details about opening and mounting the enclosure, and
explains how to install cables and PCBs. Instructions for labeling, closing and
sealing the enclosure are also provided.
BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR INTERCONNECTING
WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY (30)
SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE
PRECAUTIONS COULD RESULT IN DAMAGE TO OR DESTRUCTION OF THE EQUIPMENT AND/OR BODILY HARM.
screwdriver to loosen the sealing screw, indicated in Figure 2-1.
Figure 2-1: IND231 Screws (left) and IND236 Sealing Screw (right)
insert screwdriver to loosen the sealing screw, indicated in Figure 2-2. Note that the screws are
captive, and remain attached to the rear cover of the IND236.
CAUTION
METTLER TOLEDO IND231/IND236 Installation Manual
2-1