3. Fold the shield wire back over the outer cover of the cable.
4. Place the grounding clip supplied with the terminal over the exposed shield wire (see Figure
5. Connect the grounding clip to the enclosure using the enclosure hinge attachment screw, as
6. Place the ferrite over the load cell cable as indicated.
7. Attach the load cell wires to the connector on the motherboard. Refer to the wire color code of
30094803
|
00
|
08/2013
Figure 2-26: Load Cell Cable with Shield Wire Folded Back
2-26).
Figure 2-27: Load Cell Cable Entering IND236 Enclosure, with Chassis Ground Clip
shown in Figure 2-28 (left).
Figure 2-28: Analog Load Cell Cable Ferrite Installation and Chassis Grounding, IND236
the load cell being connected, and Figure 2-19.