3. Fold the shield wire back over the outer cover of the cable.
4. The grounding clip supplied with the terminal is used to connect the exposed shield (Figure
5. The ferrite supplied with the terminal is placed over the end of the cable, and the wires
30094803
|
00
|
08/2013
Figure 2-22: Load Cell Cable with Shield Wire Folded Back
2-22) to the metal mounting plate using the enclosure hinge wire attachment screw, as shown
in Figure 2-23.
connected to the load cell connector. Refer to the wire color code of the load cell being
connected, and Figure 2-19.
Figure 2-23: Analog Load Cell Cable Ferrite Installation and Chassis Grounding, IND231
METTLER TOLEDO IND231/IND236 Installation Manual
2-13