3.2.2.
Harsh Enclosure
The harsh enclosure is made of stainless steel with a front panel angle of approximately 38
degrees. The harsh enclosure is designed to rest on a flat surface such as a table or desk top, or it
can be mounted to a vertical surface with optional mounting brackets.
3.2.2.1.
Desktop Mounting
When the IND560x terminal will be placed on a flat surface, the four rubber feet included with the
terminal should be adhered to the bottom of the enclosure to prevent sliding. Locate the four rubber
feet from a kit of parts supplied with the IND560x, remove them from their protective paper, and
press one foot onto each corner of the bottom of the enclosure, as shown in Figure 3-9.
3-6
METTLER TOLEDO IND560x Installation Manual
By design, the rear cover plate can produce outward force on the (extended) retaining brackets,
increasing the overall stiffness of the enclosure structure and providing additional assurance
that the panel mount unit will remain securely in place. Because of this, after the Allen-head
screws have been tightened, and the unit secured into place, the rear cover plate of the
IND560x panel-mount unit may be difficult to remove and replace during servicing. If this
occurs, a slight loosening of the Allen-head screws should allow for easier removal and
replacement of the rear cover. Remember to retighten the set screws after replacing the rear
cover.
Inspect the gasket for tearing, deterioration and damage each time the Harsh enclosure is
opened. Replace the gasket if it is brittle or damaged.
Set screws
Figure 3-8: Clamping Bracket Installed
64061929
|
11
|