The 2 mV/3 mV jumper settings are described in Table 3-22.
3.9.3.
IND560x-PAB Jumper
The IND560x main board has a jumper, W3, which is specific to use of the IND560x-PAB. Both
W3 and W2 must be set, in order to use the IND560x-PAB. Figure 3-49 shows the position of
these jumpers, near the Fiber Optic and Current Loop connector.
•
•
Table 3-23 describes W3 and W2 jumper settings as they apply to the IND560x-PAB.
3-40
METTLER TOLEDO IND560x Installation Manual
Figure 3-48: Millivolt Jumper Location (left) and Positions (right)
When removing the W1 jumper for 3mV/V operation, reposition it on just one of the pins as
shown at lower right in Figure 3-48.
Table 3-22: W1 Millivolt Jumper Description
Position
Closed
Sets the connected scale's load cell setting to 2 mV/V.
Open
Sets the connected scale's load cell setting to 3 mV/V
When the IND560x-PAB is installed, both the W3 jumper and the W2 jumper must be set to ON
(closed).
When the either the PSUx or APS768x power supply is used with the IND560x, both W3 and
W2 should be set to OFF (open). In this case, the W2 jumper will return to its function,
controlling the operation of the keypad On/Off button.