Environmental
Protection
Storage
Environment
Operating
Environment
Hazardous Areas
Power
Display
Weight Display
Scale Types
Number of Analog
Cells
Number of Scales
Analog/Digital
Update Rates
1-10
METTLER TOLEDO IND560x Installation Manual
IND560x Specifications
Panel Mount front panel sealing provides IP65 (comparable to TYPE 4 and TYPE 12
protection) - certified by DEKRA EXAM GmbH.
Harsh Environment enclosure meets IP65 requirements
certified by DEKRA EXAM GmbH, and IP68 and 69K certified by MTLS.
The terminal can be stored at temperatures ranging from –20°C to 60° C (–4° to 140°
F) at 10 to 95% relative humidity, non-condensing.
The terminal can be operated at temperatures ranging from −10° to 40° C (14° to 104°
F) at 10% to 95% relative humidity non-condensing.
The IND560x is approved for use in Division 1 and Zone 1/21 2 GD hazardous areas.
Note that not all models in the IND560 terminal family can be operated in areas
classified as Hazardous because of the combustible or explosive atmospheres in those
areas. Before installation, check the data plate of the terminal to confirm it is marked
with the approval for the environment in which it will be used.
All models of the IND560x can operate with the PSUx or APS768x intrinsically safe
power supplies. The PSUx and APS768x both provide several intrinsically safe
voltages to the IND560x for its correct operation. The PSUx and APS768x are available
in either a 120 VAC or a 230 VAC version.
The IND560x harsh/analog model with the IND560x-PAB option installed operates with
the METTLER TOLEDO intrinsically safe, external NiMH battery pack
(P/Ns 64060627 & 64060625).
128 × 64 dot-matrix graphic LCD display with LED backlight. 21 mm weight digits.
Display Update Rate: 10/second
Displayed resolution of 100,000 counts for analog load cell scales.
Display resolution for high-precision IDNet and SICSpro bases is determined by the
specific base used.
Analog load cells or High-Precision IDNet, Kx Line (T-Brick Ex T4 type) or High-
Precision SICSpro, PBK9/PFK9 Category 2 bases (MPXI load cells) or Point Ex module.
SICSpro High-precision platforms that include Advanced Setup Mode. Update rate
approximately 50 Hz.
NOT COMPATIBLE WITH THE OLDER, PIK EX MODULE.
Four 350-ohm load cells (87 ohm minimum resistance), 2 or 3 mv/V
Interface for one analog or one IDNet scale or one SICSpro scale
Internal: Analog: >366 Hz; IDNet: determined by base; SICSpro: 50 Hz for bases that
include Advanced Setup Mode
Target Comparison: up to 50 Hz;
PLC Interface: up to 20 Hz
64061929
|
11
|
07/2017