A barrier that has been tested for correct operation with the IND560x, and can be connected directly
to COM1, is:
•
3.4.
Wiring Connections for Internal Options
Options that install internally to the IND560x terminal and require external wiring include the
following:
•
•
•
•
3.4.1.
IND560x-PAB
The IND560x-PAB is an optional PCB installed onto the main board of an IND560x harsh/analog
terminal. It enables this model of the IND560x to be powered by the METTLER TOLEDO external
NiMH battery pack (P/N 64060625). The external NiMH battery (P/N 64060625) is the only
battery pack approved for use with the IND560x.
This battery pack is available as a kit (P/N 64060627), which includes the NiMH battery assembly
and a 3m (9.8 ft.) connection cable. The cable is constructed with a twist lock on one end, two
3-20
METTLER TOLEDO IND560x Installation Manual
MTL7761Pac (FM and ATEX approvals)
IND560x-PAB (Power Adapter Board)
Discrete I/O
Intrinsically safe Current Loop interface
Fiber Optic interface
IMPORTANT: The IND560x-PAB option will only work with the harsh, analog version of the
IND560x terminal. The IND560x-PAB physically cannot be installed into panel mount
terminals, and will not satisfy the power needs of the IDNet interface.
IMPORTANT: The IND560x-PAB option will only work with a specific revision of the
IND560x analog main board. The correct version of analog main board can be identified
by the presence of the W3 jumper, located as shown in Figure 3-18, and indicated in
Figure 3-25. If the W3 jumper is not present on the main board, then that main board is
not compatible with the IND560x-PAB.
Figure 3-25: W3 Jumper on IND560x Analog Main Board
64061929
|
11
|
07/2017