IND560 Installation Manual
Analog Load Cell Sealing
When using analog load cells, remove the backing paper from the paper seal
included in the sealing kit and place it over the load cell terminal strip connector so
that it covers the heads of the terminal screws and also secures the connector to
the plastic plate of the IND560, as shown in Figure 2-64.
IDNet Connector Sealing
1. When using IDNet bases, the IDNet connector board must be removed from
the rear panel in order to apply the paper seal. Remove the backing paper from
the two paper seals included in the sealing kit and place the seals over both
ends of the internal IDNet harness in the IND560 as shown in Figure 2-65.
2. For IDNet bases in Europe, the plastic IDNet Sealing Kit (Part number
22000386) must be used to secure the IDNet cable to the IND560 terminal.
2-38
Figure 2-63: W1 Millivolt Jumper – Unsealed (left) and Sealed (right)
Figure 2-64: Analog Load Cell Paper Seal
Figure 2-65: IDNet Paper Seals