Information on 3A conformity

Important
Only process connections which comply with "3A Sanitary Standard 63" may be used.
The device must only be installed vertically with the terminal box or transmitter housing pointing
upward. The "bracket mounting (1)" option no longer applies.
Fig. 5
If concentric reducers are installed on the device, it must be mounted in a vertical position.
Refer to the section 3.6.9 "Installation in pipelines with larger nominal diameter".
Fig. 6
Please ensure that the leakage hole of the process connection is located at the deepest point of
the installed device.
Fig. 7
1 Leakage hole
D184B141U03