Guidelines and manufacturer's declaration – electromagnetic immunity
The UNIDRIVE
®
S III ENT model 40 7016 20-1 is intended for use in the electromagnetic environment specified below. The user of the UNIDRIVE
model 40 7016 20-1 should ensure that it is used in such an environment.
Interference immunity
EN/IEC 60601 test level
tests
Electrostatic discharge (ESD)
± 6 kV Contact discharge
acc. to IEC 61000-4-2
± 8 kV Air discharge
Rapid transient electric inter-
± 2 kV for power lines
ferences/bursts acc. to IEC
± 1 kV for input and
61000-4-4
output lines
Surges
± 1 kV Differential mode
acc. to IEC 61000-4-5
± 2 kV Common mode
Voltage dips, short interrup-
<5% U
T
tions and fluctuations
(>95% dip of the U
of the supply voltage
for ½ cycle
acc. to IEC 61000-4-11
40% U
T
(60 % dip of the U
for 5 cycles
70% U
T
(30% dip of the U
for 25 cycles
<5% U
T
(>95% dip of the U
for 5 seconds
Magnetic field at the supply
frequency
3 A/m
(50/60 Hz)
acc. to IEC 61000-4-8
Appendix
Table 202
Conformity level
6 kV Contact discharge
8 kV Air discharge
2 kV/1 kV
2 kV/1 kV
*
0 V, 10 ms
)
T
92 V, 100 ms
)
T
168 V, 500 ms
)
T
0 V, 5 sec
)
T
3 A/m
Anexo
Electromagnetic environment – guidance
Floors should be made of wood or concrete or provided
with ceramic tiles. If floors are covered with synthetic
material, the relative humidity must be at least 30%.
The quality of the supply voltage should be that of a typi-
cal business or hospital environment.
The quality of the supply voltage should be that of a typi-
cal business or hospital environment.
The quality of the supply voltage should be that of a typi-
cal business or hospital environment.
If the user of the device requests continued functioning
even when interruptions of the power supply occur, it is
recommended to supply the device from a power supply
that is free of interruptions.
The magnetic fields at the mains frequency should be at
the values typical of business or hospital environments.
®