The following applies to all modes:
The value last displayed will be saved when you quit the corresponding mode.
Inputting the password (mode 2)
Once the password has been input, modes 4 to 12 can be accessed. The password
is '99'.
Use the M button to set mode 2
Use the
If the password is correct, the other modes will be accessible.
If the password is incorrect, the unit will jump back to mode 1
Password '105' allows you to set or recalibrate the output current (modes 3a and 3b);
see Chapter 5, p. 59.
Password '110' allows you to reset to the default setting;
see Chapter 5, p. 59.
Selecting the unit of measurement (mode 4)
Once the correct password (99) has been input, use the M button to go to mode 4.
The units of measurement are displayed as follows:
bar, mbar, mmH
torr, atm
The parameters in modes 5, 6, 7, 8 and 9 are automatically converted to the new unit.
Procedure:
Use the
or
M button, and this will take you to the next mode (mode 5)
Setting the pressure measurement range (modes 5 and 6)
Upon delivery of the unit the pressure measurement range is set to the maximum
or desired range of measurement. The maximum measurement ranges are 2 bar,
10 bar, 50 bar, 200 bar or 400 bar (shown on the nameplate).
SITRANS P, ZD series
A5E00249308-01
button or the
button to select the password
O, inH
O, ftH
O, mmHg, inHg, psi, Pa, kPa, MPa, mH
2
2
2
button to select the unit of measurement, confirm by pressing the
Technical description
2
O, kg/cm
,
2
45
Tabla de contenido
loading

Tabla de contenido