Reacting to malfunction messages
Reacting to malfunction messages
Characteristic
The TORSIOTRONIC
switches off automatically
when not in use.
The TORSIOTRONIC
switches off when loaded.
The charge indicator lights
up red.
The TORSIOTRONIC
not trigger in "Click" function
mode.
Communication between the
TORSIOTRONIC
and the
®
SENSOMASTER software is
disturbed.
Display message:
Memory full!
Display message:
TARE
The display does not
disappear after 90 s at the
latest.
40
Cause
A time is set for automatic shut-off
®
when the TORSIOTRONIC
in use.
The rechargeable battery is empty.
®
The rechargeable battery is empty.
does
Electromechanical triggering is
®
defective.
The PC cannot establish a
connection to the
TORSIOTRONIC
The fastening data memory is full.
The TORSIOTRONIC
loaded during taring or the
measuring element has been
damaged due to overloading.
Remedy
The function can be switched off or
is not
the set time can be changed, see
®
page 32.
Exchange the rechargeable battery
for a full one of the same type, see
page 41.
Charge the rechargeable battery.
Exchange the rechargeable battery
for a full one of the same type, see
page 41.
Charge the rechargeable battery.
Send the TORSIOTRONIC
service department.
Check that the cable connector is
firmly seated.
.
Restart the SENSOMASTER
®
software and the
TORSIOTRONIC
Send the TORSIOTRONIC
service department.
Start the SENSOMASTER software
on the PC.
Connect the TORSIOTRONIC
the PC.
Switch the TORSIOTRONIC
The data are transferred to the PC
using the SENSOMASTER
software.
Delete the data in the
TORSIOTRONIC
is being
Relieve the TORSIOTRONIC
®
The error message disappears and
the TORSIOTRONIC
If the error message remains, send
the TORSIOTRONIC
service department.
to the
®
.
®
to the
®
to
®
on.
®
.
®
.
®
is tared.
®
to the
®
loading