Preparing for operation
When the TORSIOTRONIC
switches off after a specified period of time. The
shut-off time can be defined using the
SENSOMASTER software, see page 32.
To switch the TORSIOTRONIC
HOME button for longer than five seconds.
Three short audible signals are heard.
The TORSIOTRONIC
®
Taring
To ensure precise measurement, the
TORSIOTRONIC
must be tared for each fastening
®
process.
WARNING
Risk of injury due to incorrect
measured values caused by improper
taring.
Make sure that the
TORSIOTRONIC
loaded during the taring process.
The TORSIOTRONIC
each time when it is switched on.
Automatic taring on switching on
Place the switched off TORSIOTRONIC
flat surface so that the display points upwards.
Make sure that no forces are acting on the
TORSIOTRONIC
.
®
Keep the TORSIOTRONIC
the entire taring process.
To switch the TORSIOTRONIC
button for approx. one second.
The
symbol and TARE are shown on the
display. The TORSIOTRONIC
seconds.
22
is not in use, it
®
off, press the
®
is switched off.
is not moved or
®
is automatically tared
®
onto a
®
steady throughout
®
on, press any
®
is tared within a few
®
The completed taring process can be recognised by
the
symbol (taring process ended) on the
display.
Manual taring
Manual taring is necessary if, for example, a value
which is not zero is shown on the display after
fastening processes.
Place the switched on TORSIOTRONIC
flat surface so that the display points upwards.
Make sure that no forces are acting on the
TORSIOTRONIC
.
®
Keep the TORSIOTRONIC
the entire taring process.
Press the TARE button for longer than two
seconds.
The
symbol and TARE are shown on the
display. The TORSIOTRONIC
seconds.
The completed taring process can be recognised by
the
symbol (taring process ended) on the
display.
onto a
®
steady throughout
®
is tared within a few
®