NOTICE
Potential damage to components
► Only use the prescribed processes and the cleaning and
disinfection agents specified for cleaning and disinfecting.
Other agents, methods, dosages and contact times may
damage the components.
For information on suitable cleaning agents
and disinfectants and their specifications, see
document 9100081 at www.draeger.com/IFU.
6.2.1 Cleaning and disinfecting the device
1. Dismantle the device:
a. Separate breathing hose, facepiece and blower unit
from each other.
b. Disconnect the carrying system from the blower unit.
c. If available, dismantle any accessories (e.g. hose and
device sleeves).
d. Dismantle the splash guard cover and filter (see
"Replacing the filter", page 24).
2. Clean the facepiece according to the appropriate
Instructions for Use.
3. Clean the breathing hose and carrying system:
a. Prepare a cleaning solution containing water and a
cleaning agent.
b. Clean all parts with the cleaning solution using a soft
cloth.
c. Thoroughly rinse all parts under running water.
d. Prepare a disinfectant bath of water and a disinfecting
agent.
e. Place all parts to be disinfected into the disinfectant
bath.
f. Thoroughly rinse all parts under running water.
g. Dry all parts in the air or in a drying cabinet
(temperature: max. +60 °C). Do not expose to direct
sunlight.
4. Clean and disinfect blower unit and splash guard cover
using disinfectant cloths.
In cases of strong contamination, the blower unit can be
rinsed under running water as follows.
1. Make sure the rechargeable battery remains inserted.
Water must not enter the battery compartment.
2. Close suction inlet and tube connection with plug
(available as accessories).
6.3
Maintenance work
6.3.1 Visual inspection
Check all parts thoroughly and replace damaged parts if
necessary. In particular, check the filter sealing surface and
O-rings of the blower unit for damage (e.g. scratches) or
contamination.
Instructions for use
|
Dräger X-plore
®
8000
6.3.2 Replacing or charging the rechargeable
battery
WARNING
Explosion, fire or chemical hazard!
► Do not remove, insert or charge rechargeable batteries in
potentially explosive or flammable environments.
► Keep rechargeable batteries away from sources of heat.
► Do not short-circuit the rechargeable battery contacts.
► Only use recommended rechargeable batteries.
Removing the rechargeable battery:
1. Fold up carrying system if necessary.
2. Push battery lock button. Ensure that the rechargeable
battery does not fall down.
3. Remove rechargeable battery.
Inserting the rechargeable battery:
1. Fold up carrying system if necessary.
2. Position the two lower connection points of the
rechargeable battery at an angle in the battery
compartment sockets.
3. Fold the rechargeable battery in until it snaps audibly into
place.
Always disconnect the charger from the power supply if not
in use.
Always disconnect the charger from the power supply if not
in use.
The standard rechargeable battery (EX) and the long-life
rechargeable battery (EX) may only be charged using the
Dräger X-plore 8000 standard charger or the Dräger X-plore
8000 multiple charger.
Charging the battery:
1. Check to make sure that voltage of mains supply is
correct. The operational voltage of the power supply unit
must match the mains supply voltage.
2. Connect charger to power supply unit.
3. Connect the power supply unit to the mains supply.
4. First position the rechargeable battery at an angle in the
charger and then fold it in until it snaps audibly into place.
5. Wait for the end of the charging process.
6. When the rechargeable battery is fully charged, push the
battery lock button and remove the battery.
7. Disconnect the power supply unit and charger from the
mains supply.
|
en
Maintenance
23