Operation
7.5.2 Opening menus
● To open the Information menu:
a. Tap and hold
for 1 second in measuring mode.
● To open the Calibration menu:
a. Tap and hold
for 3 seconds in measuring mode.
b. Select Enter password.
c. Enter the corresponding password and confirm with
[OK].
● To access all menus:
a. Tap and hold
for 3 seconds in measuring mode.
b. Select Enter password.
c. Enter the corresponding password and confirm with
[OK].
7.5.3 Passwords
Calibration
Permits access to the Information menu
and the Calibration menu (adjustment
PWD
settings)
Default value: _ _ _ 1
Settings PWD
Permits access to all configurations and
menus
Default value: _ _ _ 2
Information on changing the passwords can be found in 11.1.
7.5.4 The Information menu
In the Information menu, the following information can be
called up:
Instrument
Warnings
Warning messages in plain text, including
the corresponding number.
If there are several warning messages, this
will be shown accordingly (e.g. with 1/3 =
screen 1 of 3).
Faults
Error messages in plain text, including the
corresponding number.
If there are several errors, this will be
shown accordingly (e.g. with 1/3 = screen 1
of 3).
Device flag With this function, the warning and error
codes are shown in table format. If the
value 00 is shown for all codes, then there
are no warning or error messages present.
An overview of the installed hardware mod-
Modules
ules, as well as the software version of the
SIOS-specified EC sensor interface, will be
shown. For more detailed information,
select the corresponding module.
= installed modules
= modules not installed
Sensor
Last cal.
Adjustment date, unit, gas and concentra-
date
tion of the last adjustment are shown.
18
Shows the date of the adjustment that is
Next cal.
date
due next.
The sensor vitality is shown in %. Dräger
1)
Vitality
recommends replacing the sensor when
vitality is < 25 %.
The current and highest sensor tempera-
Sensor
tempera-
ture will be shown.
1)
ture
The sub-menu shows the number of days
Sen-
sorAlive
left for the expected service life and the
vitality in %.
Data logger
Logger
Data logger status display (active or inac-
tive)
status
Display of progress over last 15 minutes in
Graph
a time/concentration graph.
1) The function is only available with the diagnostic dongle.