Number
Cause
167, 170
Adjustment inter-
val expired.
Only for Polytron 8100 EC
Number
Cause
163
Low EC sensor
vitality
171
Negative reading.
Value below range
minimum.
Only for Polytron 8100 EC with Dräger SensorAlive
Number
Cause
320
Communication
with the sensor
failed. Diagnosis
and gas generator
no longer work.
321
Evaluation of the
sensor data failed.
The sensor cannot
perform or evalu-
ate a LifeTest (pre-
conditions are not
met).
322
Sensor data set is
faulty.
323
Sensor data set is
incompatible.
324
Sensor sensitivity
is less than 50 %
(calculated value).
325
Sensor sensitivity
is less than 40 %
(calculated value).
326
Sensor cannot be
adjusted during a
LifeTest and for a
certain period
afterwards.
327, 330,
Gas generator is
344, 347
not working.
331
Gas generation
failed.
332
Sensor is in the
safe state. No Life-
Tests are per-
formed (group
warning).
Instructions for use
|
Dräger Polytron
Remedy
Readjust the instrument.
Remedy
Replace sensor.
Recalibrate zero point.
Remedy
Remove the sensor, do not
use it for approx. 1 day and
reinsert it.
Check the firmware version
and have it updated by
DrägerService if necessary.
Replace the sensor.
Change the time of the
LifeTest or set the repeat
profile to "Flexible" (see
"Dräger SensorAlive set-
tings", page 36). If the
warning occurs again, con-
tact DrägerService.
Replace the sensor.
Contact DrägerService.
Adjust the sensor and pre-
pare for sensor replace-
ment (e.g., order the spare
part).
Adjust or replace the sen-
sor.
To be able to perform a
scheduled adjustment, the
LifeTest can be temporarily
inhibited in PolySoft via the
Inhibit mode.
Adjust or replace the sen-
sor.
Adjust the sensor.
Determine the cause and
rectify it.
®
8000 Series
Number
Cause
333
Real time clock of
sensor is not work-
ing correctly.
334
Sensor synchroni-
zation not com-
plete.
336
Internal sensor
checks return
faulty results.
337
Transmitter firm-
ware must be
updated.
340
Sensor is synchro-
nized after the sys-
tem is restarted.
342
Temperature differ-
ence between the
two temperature
measurement
points is too great.
347
Configuration
failed due to an
interruption in the
communication
between the trans-
mitter and the sen-
sor.
350
Gas generator is in
the start-up phase.
352
Sensor self-test
failed.
353
Sensor is tempo-
rarily in Inhibit
mode. No LifeTests
are performed
during this time.
354
Error during sys-
tem start.
355
Sensor firmware is
incompatible with
the hardware.
356
Hardware data
faulty.
357
Sensor was set to
the safe state man-
ually, e.g., to per-
form a firmware
update.
Troubleshooting
|
en
Remedy
Set the clock.
Wait until synchronization
is complete. Depending on
the condition of the sensor,
this state may last for sev-
eral minutes. Have the
firmware updated by
DrägerService if necessary.
Restart the system.
Contact DrägerService.
Wait until synchronization
is complete.
Wait and check if the
LifeTest is evaluated, or
replace the sensor.
Reconfigure the sensor
and check the configura-
tion.
Wait until the start-up
phase has ended. During
the start-up phase, the
error detection and clog-
ging detection functions are
not active.
Replace the sensor or con-
tact DrägerService.
Inhibit mode can be exited
manually via PolySoft or is
exited automatically.
Perform manual restart.
Have the firmware updated
by DrägerService.
Perform a restart or replace
the sensor.
To exit the mode, restart
the sensor.