Installation and starting
• Unpack the appliance following the instructions shown on the package.
• Dispose of the package in compliance with regulations.
• Verify that the appliance has not been damaged during transport.
• Do not connect damaged appliances to the mains.
• Do not use extension leads, multiple plugs or sockets.
• Ascertain that the feeding line is adequate to feed the machine.
• Ascertain that the appliance (or Blok-Jet) is installed in a clean spot, far from heat
sources and from depots of contaminating substances, and that it aspirates clean
air, free from dust, gases, and humidity.
• Any machine installation must be carried out by a specialist, duly trained and equipped
with necessary tools.
The installer must refer to the appliance manual, carry out the machine fi nal test and
instruct users on use and routine maintenance of new not used and therefore not yet
contaminated machines.
• Before starting the compressor and/or the aspirator ascertain that piping is clean; heavy
debris could damage the appliances.
• Danger signals and warnings (light and beepers) such as failing-machine or overheating
warning must be removed from technical room and installed in room under frequent
control to be signalled by a sound or visual warning.
• Air exhausted from aspirators should be fi ltered, through a suitable certifi ed absolute
fi lter and then conveyed outside.
• After fi nishing the installation, carry out functional and regulation tests; check the rota-
tion sense of motors, the mains tension, and electrical absorption.
• Check the functioning of drainages, aspiration fl ow rate and head, compressor charging
time, functioning of silica-gel regenerating device, and running of ventilators.
• Make sure that periodical checks are carried out on equipments: such an inspection is
not only a means to protect the surgery from forced stops and to check its operation
conditions, but it is also a proper way to prevent accidents.
• Our updated manual are in our web site www.cattani.it. We recommend to consult
them especially for safety up-dating.
Install and connect the aspirator as per wiring diagram, (Fig. 8) select the ON position
on the switch, which will light up. At this point aspiration will start.
The control panel is fi tted with a keypad which allows access to the program, to regulate
or to modify the head value and with a display showing the current working parameters
of the system.
It is allowed to enter the program to interrogate the operation log, especially those con-
cerning alarms. In case of alarms it is also possible to place a remote fl ashing signal,
in the reception offi ce or in another room.
(wiring diagram Fig. 8, page 77)