3     Features and Functions
Power On
Configures the instrument's power-on state and power-on message.
Softkey
Power On
Last
User
Defined
Factory
Defaults
Power On
Message
Display Options
Configures the instrument's display.
You can enable or disable the display, adjust the brightness (10 to 100%), pick a color scheme, and enable or disable
the screen saver. If you turn off the display, press any front panel key to turn it on again. By default, the screen saver
increases display life by dimming the display during long periods of inactivity. You may disable this screen saver from
the front panel only. The display is enabled when power is cycled, after an instrument reset (*RST), or when you
return to local (front panel) operation. Press the [Local] key ([Save Recall] key) to return to the local state.
DMM On/Off
Enables or disables the internal DMM. The
disabled, the DMM Digitize option will not be available when selecting the data acquisition mode.
Language
Selects the local language for front panel use.
All messages, context-sensitive help, and help topics appear in the selected language. The menu softkey labels are
not translated.
76
Description
At power up, the instrument loads the previous state it was in during power down.
At power up, the instrument loads up a user-selected state file.
At power up, the instrument loads up the factory default state.
Power On Message sets a message that is displayed when the instrument powers up and when you press
[Home] > Help > About . Use the knob or the front panel arrow keys and [Select] key to select the letters. Then
press
Done to exit and save the message. The message will appear as shown below when you turn the
instrument on or press
[Home] > Help > About .
annunciator will display when internal DMM is disabled. Once