3     Features and Functions
Cursors in the histogram are specified as bin numbers and display the range of values covered by those bins, the
count and the percentage of the total. The total count and percentage of total, as well as the measurement range
covered, between the cursor bins, are also displayed.
In the example below, cursor B1 (violet vertical lines) is positioned on bin number 30 and cursor B2 (green vertical
lines) is positioned on bin number 50 (bin number shown above the B2 softkey). The bin information for cursor B1 is
shown in the violet box and the bin information for cursor B2 is shown in the green box.
Based on the example above, the information for cursor B1 can be interpreted as follows:
– 0.126m to 0.120m - Range of DCV measurement values for this bin.
– #443 - Number of samples in this bin.
– 1.19% - Percentage of the total number of samples in this bin.
The data between the B1 and B2 cursors, including the data in the B1 and B2 bins, is shown to the right of the violet
and green boxes. In the graphic above:
– Span: 0.126m - The measurement range covered from B1 to B2.
– #15.6k - The number of samples from B1 to B2.
– 42.1% - The percentage of the total number of samples from B1 to B2.