3 Features and Functions
Scanning with External Instruments
You can use the DAQ970A/DAQ973A for signal routing or control applications. If you install a multiplexer plug-in
module, you can use the DAQ970A/DAQ973A for scanning with an external instrument. You can connect an external
instrument (such as a DMM) to the multiplexer COM terminal.
To control scanning with an external instrument, two control lines are provided. When the DAQ970A/DAQ973A and
the external instrument are properly configured, you can synchronize a scan sequence between the two.
Keysight DAQ970A/DAQ973A User's Guide
57