Copy/paste from multiple channels to multiple channels (many-to-many)
For example, to copy/paste the measurement configuration from channels 01 through 04 to channels 05 through 08:
1. Press [Channel] and use the knob or the front panel arrow keys to configure the source channels as below:
– Channel 01 and 02 (DCV with Autorange)
– Channel 03 and 04 (2-wire resistance with the range of 100 Ω)
2. Press [Copy] > Source Chan(s). Use the knob or the front panel arrow keys to determine the Start Channel as
channel 01 while End Channel as channel 04. Press Done to exit this menu.
3. Press Dest. Chan(s) to select the destination channel for the paste operation. The instrument will automatically
select the subsequent channel as default choice of destination channels. In this example, use the knob or the
front panel arrow keys to determine both Start Channel as channel 05.
End Channel softkey is unavailable in this option, instrument will automatically perform the copy/paste oper-
ation into the next N subsequent destination channels based on the N selected of source channels. For example,
if you set the Start Channel as channel 06, destination channels selected will be channels 06 through 09. Press
Done to proceed.
4. Press Copy Now. The measurement configuration from channels 01 through 04 will copy to channels 06 through
09. "COMPLETED" message will prompt out when the operation is completed. Press Copy Now again to
repeatedly copy the same configuration to the new selected channels (channels 09 through 11, and so on).
Keysight DAQ970A/DAQ973A User's Guide
3 Features and Functions
169