Bar Meter
The bar meter adds a moving bar below the standard Number display.
Softkey
Scale
Default
Manual
Limits
Peak On/Off
Clear Peaks
Clear Mon Data
Statistics
Hide/Show
Keysight DAQ970A/DAQ973A User's Guide
Description
Default sets the scale to the measurement range.
Manual allows you to configures the scale either as High and Low values or as a Span around a Center
value. For example, a scale that goes from a
Center of 250 Ω with a Span of 1500 Ω.
a
Limits set the scale to the alarm limits.
This softkey is only available when you configure alarm on the selected channel
by pressing the [Alarm] key on the front panel.
Press the
Peak softkey to toggle between enabling ( On ) or disabling ( Off ) the measurement of the
Minimum, Maximum, and Peak to Peak values acquired within the measurement aperture.
Press the
Clear Peaks softkey to reset the reading of peak values. This is done by clearing the
accumulated history of peak-to-peak function.
The Clear Peaks softkey will only be displayed when Peak is set to On.
Press the
Clear Mon Data softkey to clear the statistics, trend chart, and histogram on the monitor
channel.
The Clear Mon Data softkey will only be displayed when Statistics is set to Show.
Press the
Statistics softkey to toggle between Hide or Show the measurement statistics for the monitor
channel. Monitor channel statistics are cleared when monitor mode is turned on, when the monitor
channel changes, or when the monitor channel's configuration is changed.
3 Features and Functions
Low of -500 Ω to a High of 1000 Ω could also be specified as