5.8 Autoclaving
The following applies only to model PG43SA-S with option "autoclav-
able version", recognizable on the dial printing as "AUTOCLAVABLE".
Observe the permissible conditions of autoclaving, see chapter
9.3 "Permissible temperature"
During the autoclaving process, no weight-loading must be exert-
ed via the case and the instrument must not lie on its window.
After the first autoclaving cycles, the window slightly arches
outward. This has no effect on the function of the instrument.
Depending on the autoclaving process, whitish residues can build
up on the window and pointer. This has no effect on the function of
the instrument.
In case of a zero offset, carry out a zero point setting. See chapter
5.7 "External zero point setting".

6. Faults

Following any failure or shut-down of the plant all instruments should
be checked and, if necessary, replaced before recommissioning the
plant.

7. Maintenance and cleaning

7.1 Maintenance
The instruments are maintenance-free.
The indicator should be checked once or twice every year. For this
the instrument must be disconnected from the process to check
with a pressure testing device.