Storz IMAGE1 S 4U Rubina Manual De Instrucciones página 49

Ocultar thumbs Ver también para IMAGE1 S 4U Rubina:
47
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the IMAGE1 S™ 4U Rubina,
The IMAGE1 S™ 4U Rubina, OPAL1
NIR/ICG camera head is intended for use in an electromagnetic environment in which radiated RF disturbances
®
are controlled. The customer or user of the IMAGE1 S™ 4U Rubina, OPAL1
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the IMAGE1 S™ 4U Rubina, OPAL1
NIR/ICG camera head as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of structures,
objects and people.
HF surgical instrument compatibility
This equipment has been evaluated for
compatibility with high­frequency surgical
equipment. It has been tested according to
IEC 60601­2­2 Annex BB.
Information on
electromagnetic
compatibility (EMC)
Table 5
OPAL1
NIR/ICG camera head
®
NIR/ICG camera head can help prevent electromagnetic interference by
®
Separation distance d [m] as a function of the transmitter frequency
150 kHz to 80 MHz
80 MHz to 800 MHz
f
0.12
0.38
1.2
3.8
12
Indicaciones sobre
compatibilidad
electromagnética (CEM)
800 MHz to 2.5 GHz
f
h
0.12
0.38
1.2
3.8
12
®
0.23
0.73
2.3
7.3
23