IMPORTANT PARTS OF YOUR CONCENTRATOR
Please take time to familiarize yourself with your DeVilbiss oxygen concentrator before operating
1
2
A
6
Front View (Figure A)
1.
Operating instructions (LED indicator panel)
2
Power Switch
| = ON
O = OFF
3
Flow meter knob
4
Flow meter
5.
Circuit breaker – resets the unit ater electrical overload shutdown
6.
Oxygen outlet tting – oxygen is dispersed through this port
7.
Normal Oxygen (green) light (see page 9)
8.
Low Oxygen (yellow) light (see page 9)
9.
Red Service Required (red) light – when illuminated contact your DeVilbiss
provider
10 Hour meter
Accessories
Transfiller Caddy
Bubble Humidifier
There are many types o humidiers, oxygen tubing and cannulas/masks that can be used with this device. Certain humidiers and accessories may impair the device's
perormance. A mask or any nasal cannula can be used with continuous fow delivery and may be sized according to your prescription as recommended by your homecare
provider who should also give you advice on the proper usage, maintenance and cleaning
WARNING
The accessories (nasal cannula, masks, oxygen tubing, humidiers, etc.) that supply oxygen to the patient must be equipped with a means that, in case o re,
stops the propagation o re through the accessory or the saety o the patient and others. A re activated fow-stop or thermal use device, i available, should be
used with the oxygen supply accessories. These types o fow-stop devices stop the fow o oxygen to the patient in the event o re. This means o re protection
should be located as close to the patient as practicable
WARNING
When using the Transller Caddy with a Transll device, always keep the system on a fat surace. Disassemble the system prior to moving.
NOTE – A maximum o 50 eet (15 meters) o crush-proo oxygen tubing, plus 7 eet (2.1 meters) o cannula, plus a bubble humidier is allowed between the concentrator
and the patient.
NOTE – Your healthcare provider should veriy the compatibility o the oxygen concentrator and all o the parts used to connect to the patient beore use.
A-525DS
8
7
9
3
4
5
10
B
15
16
11
14
12
13
Back View (Figure B)
11 Handgrip
12 Exhaust Vents
WARNING
When the device is used under extreme operating or single fault
conditions, the exhaust air near the exhaust vents on the bottom of
the unit may exceed 41˚. Keep exposed body parts, such as hands
and eet, a minimum o 46 inches (1.2 meters) away rom the exhaust
vents to avoid the risk of burns
13
Power cord and/or IEC power connector
14 Line cord strap
15. Filter Door with venting and compartment or optional gross particle lter
16. Auxiliary Oxygen Port (Serial numbers starting with R, N, or B): Your
concentrator is equipped with an auxiliary oxygen port that can be used to ll
oxygen cylinders with an FDA-cleared cylinder lling device that is designed
to use oxygen rom a concentrator to ll a cylinder. The port is only or use
with FDA-cleared lling devices with compatible oxygen input specications.
Reer to the cylinder lling device instruction guide or the oxygen input/
output specications, connection and operating instructions.
11
12
DeVilbiss 525DD-650
Salter Labs 7600 or equivalent
EN - 7