DeVilbiss Healthcare 525 Serie Guía De Instrucción página 5

Ocultar thumbs Ver también para 525 Serie:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

WARNING
Risk of injury or damage
When the device is operated at the extremes o the environmental operating specications (i.e., maximum temperatures and humidity), and in a single ault condition,
which is a single component or performance malfunction, such as a blocked exhaust vent, blocked air intake, or an internal cooling fan failure, the temperature of the air
coming out o the exhaust vents, located on the bottom let and bottom right sides o the unit can reach temperatures capable o causing a burn injury (see stated
temperature and contact time values in the model specic tables below).
Keep exposed body parts, such as hands and eet, a minimum o 46 inches (1.2 meters) away rom the exhaust vents to avoid the risk o burns. Single ault conditions
may result in visual and audible alerts and alarms
NOTE – Under normal and single fault conditions, the concentrator releases warm air out the bottom of the unit (exhaust vents) which may discolor temperature
sensitive fooring suraces. The concentrator should not be used over fooring that is sensitive to heat staining. The Manuacturer is not responsible or fooring that
becomes discolored.
Operate the unit in a cool, dry area with good ventilation, located on a hard surface, avoid thick rugs or carpeting NEVER block the air intake or exhaust vents Keep
the unit a minimum o 12 inches (30.5 cm) away rom any wall, draperies, or any other objects that might prevent the proper fow o air in and out o your oxygen
concentrator. Proper air fow is needed to prevent overheating o the oxygen concentrator. DO NOT place the concentrator near any heat source such as hot air
registers or heaters Overheating of the oxygen concentrator may lead to low oxygen output and a risk of burns
The oxygen concentrator should be located in a well-ventilated area. DO NOT operate the unit in a closed or conned space, such as a closet, bathroom, etc. Avoid
operating the device near smoke pollutants and fumes
Under extreme environmental conditions and a single ault condition occurs, the ollowing device surace temperatures may exceed 106 °F (41 °C). See Table 1 below
or the model specic maximum temperature and sae contact guidance:
Table 1 - 525DS Series
Description
Air coming from exhaust vents located on each side
near the bottom of the unit
Oxygen outlet tting
Power Switch
LED Indicator panel
Cannula at the outlet
Table 2 - 525KS Series
Description
Air coming from exhaust vents located on each side
near the bottom of the unit
LED Indicator panel
Table 3 - 525PS Series
Description
Air coming from exhaust vents located on each side
near the bottom of the unit
Power Switch
LED Indicator panel
CAUTION
Federal (U.S.A.) law restricts this device to sale by or on the order o a physician.
It is very important to ollow your oxygen prescription. Do not increase or decrease the fow o oxygen – consult your physician.
Use o harsh chemicals (including alcohol) is not recommended. I bactericidal cleaning is required, a non-alcohol based product should be used to avoid inadvertent
damage
IMPORTANT
It is recommended that the homecare provider lock the fow control knob to prevent inadvertent adjustment. A fow setting other than prescribed may aect the patient
therapy
Do not service or clean this device while in use with a Patient
Installation o 515LF-607 low output fow meter package will cause the low fow alarm to not work and will prevent the device rom meeting the requirements o ISO-
80601-2-69:2014 Section 201 13 2 101
The Device is classied as IP21 which means it is protected against nger access to hazardous parts and protected against vertically alling water drops.
Equipment not suitable or use in the presence o a fammable anesthetic mixture with air or with oxygen or nitrous oxide.
This device contains electrical and/or electronic equipment Follow local governing ordinances and recycling plans regarding disposal of device components
A-525DS
147 0
120 7
119 0
129 5
109 1
169 0
110 5
SAVE THESE INSTRUCTIONS.
Maximum Temperature
525DS Series
°F
°C
Max safe contact time
63 9
Less than 1 minute
49 3
Less than 10 minutes
48 3
Less than 10 minutes
54 2
Less than 1 minute
42 8
Less than 10 minutes
Maximum Temperature
525KS Series
°F
°C
Max safe contact time
76 1
Less than 10 seconds
43 6
Less than 10 minutes
Maximum Temperature
525PS Series
°F
°C
142 2
61 2
106 7
41 5
108 5
42 5
Max safe contact time
Less than 1 minute
Less than 10 minutes
Less than 10 minutes
EN - 5
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

525ds serie525ks serie525ds-q serie525ks-lt serie525ps serie

Tabla de contenido