(GB-IE)
Er0
THERMAL BOILER
Er1
TANK FILLING TIME-OUT
Er2
THERMOSTOP: BOILER NOT AT
TEMPERATURE AFTER
EXTENDED WASHING
Er3
TANK FULL AFTER MANUAL
DISCHARGE
Er4
BOILER PROBE BROKEN/NOT
CONNECTED
Er5
TANK PROBE BROKEN/NOT
CONNECTED
Er6
BOILER HEATING
If the Boiler Safety Thermostat trips, the machine is switched off and
disconnects completely. The Er0 error message flashes on the TANK
display (7), and the ON/OFF (2) LED and START key (3) both flash red.
N.B.: Check and fix the cause of the Safety Thermostat error before
resetting it.
If the water in the tank does not reach the specified level after 5 minutes,
the filling valve is switched off. The Er1 error message flashes on the TANK
display (7) and the START (3) key flashes red.
To reset the alarm, press the START key (3) for 3 seconds to resume the
filling, or switch off the machine by pressing the ON/OFF key (1).
N.B.: Check and fix the cause of the failure before restarting the machine.
To restart the machine, press the ON/OF key (1) for about 2 seconds
With the Thermostop enabled (LED 10 on), if the boiler is not at the set
temperature, the washing time is extended by a maximum of 8 minutes. If
the boiler does not reach the set temperature within the extended time, the
cycle nevertheless passes to the pause and rinse phase. At the end of the
washing cycle, the Er2 error message flashes on the BOILER display (6)
and the START key (3) flashes red.
The error is reset when the boiler temperature is reached, or when the
machine is switched off by pressing the ON/OFF key (1) for 3 seconds.
N.B.: Check and fix the cause of the boiler's failure to heat up, before
switching it on again.
The discharge lasts for 3 minutes, at the end of which the tank level is
checked: if the level is high, the Er3 error message flashes on the TANK
display (7) and the START key (3) flashes red.
The error is reset by pressing the ON/OFF key (1) for 3 seconds to switch
off the machine, or by pressing the START key (3) for 3 seconds to allow
the discharge to resume from the beginning.
N.B.: Check the correct functioning of the discharge pump before restarting.
If the boiler probe short circuits or is interrupted, the Er4 error message
flashes on the BOILER display (6) and the START key (3) flashes red.
To reset the alarm, switch off the machine by pressing the ON/OFF key (1)
for three seconds.
N.B.: Check the status of the probe before switching the machine back on
again.
If the tank probe short circuits or is interrupted, the Er5 error message
flashes on the TANK display (7) and the START key (3) flashes red.
To reset the alarm, switch off the machine by pressing the ON/OFF key (1)
for 3 seconds.
N.B.: Check the status of the probe before switching the machine back on
again.
If the boiler does not reach the set temperature after 30 minutes, the Er6
error message flashes on the BOILER display (6) and the START key (3)
flashes red. All the functions remain active. To reset the alarm, switch off
the machine by pressing the ON/OFF key (1) for 3 seconds.
N.B.: Check the correct functioning of the boiler resistance before switching
it back on again.
53