WARNING
Risk of deviating measured values during flow
measurement
Applies to Cato, Cicero, and Narkomed 6000:
Manual ventilation may result in deviating measured
values.
Replacing/Snapping into place
The disposable CO
absorber must be replaced as soon as one of
2
the following conditions applies:
–
The inspiratory CO
concentration has reached approx.
2
5 mmHg.
–
The soda lime in the absorber has turned 2/3 violet.
–
The disposable CO
absorber has been used for 4 weeks or
2
longer
NOTE
The color indicator can regenerate slowly and the soda lime may
revert to white, but its absorption capacity is nevertheless spent.
The color change is only an approximate indicator; the CO
monitoring is the decisive factor.
Always remove spent disposable CO
and dispose of immediately.
Replace the disposable CO
2
continuous CO
absorption in the breathing system.
2
1
Press the button: the disposable CO
2
Remove the disposable CO
3
Before fitting into place, shake the new disposable CO
absorber by turning it upside down several times in order to
loosen up the soda lime.
Make sure to fit the disposable CO
correctly (pictograms must face the user).
4
Remove the red cap.
5
Slide the disposable CO
absorber onto the mounting as far as
2
possible.
10
absorbers from the device
2
absorber quickly to ensure
absorber swings out.
2
absorber from the mounting.
2
2
absorber into place
2
Instructions for use CLIC Absorber 800+ / Infinity ID CLIC Absorber 800+ / CLIC Adapter
6
Swing into the system until it snaps into place.
2
The disposable CO
CLIC adapter.
CO
bypass function
2
WARNING
Risk of patient injury
There is no CO
absorption in the breathing system if the
2
disposable CO2 absorber is swung out.
Activating the CO
2
Press the button: the disposable CO
Deactivating the CO
Swing the disposable CO
snaps into place.
absorber must be fully covered by the
2
bypass function
absorber swings out.
2
bypass function
2
absorber into the system until it
2