HighSurg 30
Parameter Level 2
Values in Parameter Level 2 can only be adjusted after the password "9403" was entered. The password cannot be changed.
Entering password:
Press "Enter", press "Select
To confirm press "Enter" for 1 second, until a tone is emitted.
Handpiece
Max. Speed
Handpiece max speed/HP 01
Handpiece max speed/HP 02
Handpiece max speed/HP 03
Handpiece max speed/HP 04
Handpiece max speed/HP 07
Handpiece max speed/HP 08
Handpiece max speed/HP 09
Handpiece max speed/HP 10
Handpiece max speed/HP 11
Handpiece max speed/HP 12
Handpiece max speed/HP 13
Handpiece max speed/HP 14
Handpiece max speed/HP 15
* Speed selection in 4 steps: 9000, 10'000, 11'000, 12'000 rpm
Pump parameters
Pump/Backwards turn mode variable
Pump/Way backwards
Pump/Speed backwards
Pump/Range 1 increment
Pump/Range 1 end
Pump/Range 2 increment
Pump/Range 2 end
Pump/Range 3 increment
Pump/Range 3 end
Pump switch mode on/off
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
All manuals and user guides at all-guides.com
▲
▼
or
", (for fast forward or backward keep key pressed).
Name of handpiece
Speed range
on the display
rpm
1 : 5
1500 – 240,000 240,000
1 : 3
900 – 150,000
1 : 2
600 – 100,000
1 : 1
300 – 80,000
Kirschner Hp
500 – 2800
Tattoo Hp
9000 – 12,000* 12,000
Cont. Shaver
300 – 6000
Osc.Shaver
300 – 5000
Jacobs Chuck
200 – 2600
Perforator
80 – 1200
Craniotome
1000 – 60,000
4 : 1
75 – 20,000
OTO-Drill Hp
300 – 16,000
Range
No/Yes
1 – 100%
10 – 50%
1 – 10%
5 – 50%
1 – 10%
10 – 90%
1 – 10%
20 – 100%
No/Yes
Factory
Definition
default
Limit the maximum speed of your
handpieces according to your own
150,000
experience.
100,000
80,000
2800
6000
5000
2600
1200
60,000
20,000
16,000
Factory
Definition
default
Yes
The pressures in the tube set vary according to
the pump speed. In "variable mode" this is
considered, to prevent of spilling when pump
is switched off in backwards mode.
25%
Specify how far the pump turns backwards
33%
Specify how fast the pump has to turn
backwards to prevent of spilling after the
switch off of the motor.
1%
Adjustment steps in section 1.
20%
Set the range where section 1 is active
5%
Adjustment steps in section 2
50%
Set the range where section 2 is active.
10%
Adjustment steps in section 3
100%
Set the range where section 3 is active.
No
No: Pump starts simultaneously with the
handpiece drive.
Yes: Pump starts separately by pressing the
foot switch.
EN
19