Shaver functions table, available for the 50,000 and 80,000 rpm motor
Name of handpiece with
transmission ratio
Continuous Shaver
Oscillating Shaver
The oscillation period for oscillating shavers is determined by the speed. If the oscillation period is set too low for
the set speed, it switches automatically to the next possible level.
This ensures that the shaver knife makes a full rotation for the speed setting.
Handpieces that don't belong to one's own assortment can be deactivated in the Configuration Menu. (refer to "7.7
Configuration Menu") That makes finding the right handpiece in the handpiece list easier, because the list of
available handpieces becomes shorter.
7.4.2
Step 2: Calibrating handpieces
To make sure the displayed parameters correspond with the actual, measurable parameters of the
handpiece or contra angle, it is recommended to calibrate each handpiece or contra angle on a regular basis.
It's a procedure as simple as it is important to guarantee safety and precision of each handpiece or contra
angle being used.
After you take care of all prior preparations such as sterilization, maintenance and care of handpieces,
device preparation and the selection of the handpiece of use, the calibration procedure can be performed.
The calibration of handpieces or contra angles guarantees accurate torque. Due to wear as well as varying
lubrication of the handpieces or contra angles, and lack of maintenance and care, the distribution of
torque can vary widely.
1. Select the correct handpiece, corresponding to the one mounted on the motor and double check on the
display for correct selection.
2. Hold motor with mounted handpiece or contra angle in your hand at safe distance to your body.
3. Press both "Speed" keys at the same time (Speed ▲ + ▼). "Testing the handpiece XX" is displayed.
4. Motor and handpiece start running and pass several speed cycles up to maximum speed.
5. After a tone is emitted the calibration is finished. The display shows the message "Handpiece is OK".
If a handpiece does not work within the parameters established by calibration even after cleaning and
lubrication, the device displays a red-illuminated error message, "Handpiece XX is faulty".
This indicates soiling, wear or a technical defect. Such handpieces must be repaired or replaced.
The torque response of all handpieces is checked during calibration.
With the 1:1 handpiece, control of the device is also adjusted to altered handpiece conditions to remain
within the torque response tolerance.
14
All manuals and user guides at all-guides.com
Speed
Display
min. rpm
Cont. Shaver
300
Osc. Shaver
300
Oscillation
Speed
period
max. rpm
from ... seconds
6000
5000
0.20
HighSurg 30 Operating Instructions, Ref. 31666, V24/16
HighSurg 30
Oscillation
period
to ... seconds
3.00