Chapter
2
The device must be installed and used in accordance with the safety regulations and
instructions for use supplied in this User Manual, for the purposes and applications
for which it is intended.
Important:
All external surfaces of the sensor, sensor cable, sensor holders and the sensor
sheaths are considered to be applied parts and are safe for normal or accidental
patient contact during use.
The DEXIS Titanium has no serviceable parts and does not require calibration. Do
not open the device to service it. All aspects of the DEXIS Titanium that are meant
to be attended to by the operator are accessible without opening the internal
components of the device. If there is a service problem, contact a qualified dealer
service representative or DEXIS Technical Support.
Important:
DEXIS Titanium is a USB compliant device and shall be used with USB compliant
cables suitable for high speed/USB 2.0 cables. Such cables are either marked "USB
2.0" or "USB Hi-Speed."
USB certified hubs or active booster cables can be used to extend the distance to
the USB host/computer. The length of the cable connection to the hub or between
hubs shall not exceed 5m.
NOTE: 3 ft. Booster Cable (1.014.1145) is available for purchase and sold
separately.
CAUTION
Using non-USB compliant cables or hubs, or exceeding the maximum
count of USB hub devices for extending the distance, can degrade the
immunity of the DEXIS Titanium sensor to electromagnetic fields or
increase the emission of electromagnetic fields from the sensor.
English
Safety and Disposal
Procedures
Any serious incident that has occurred in relation to the use of
the DEXIS Sensor should be reported to Technical Support at
1-888-275-5286.
Portable/mobile radio frequency communications equipment
can affect the function of the DEXIS Titanium sensor as well as
any other electronic medical equipment.
2-1