DEXIS Titanium
Environmental Conditions
USAGE
DEXIS Titanium sensor is not
suitable to be operated in
oxygen rich and/or explosive
environments
TRANSPORTATION AND
STORAGE
Transport in the supplied
protective package
EU Classification
Protection against water/matter – IP 68

Compatibility with Radiographic Generators

DEXIS Titanium is generally compatible with any dental X-ray unit and generator
capable of supplying the required range of exposure times and doses.
Follow the instructions of the X-ray generator to set the desired dose range.
The radiation of X-ray tubes is controlled by the settings of:
Some controls allow for the modification of all of the above, some have fixed settings
for current and voltage.
Distance of X-ray Source to Sensor
There is a correlation between the distance of the X-ray source to the sensor and
the dose received by the DEXIS Titanium sensor. The radiation that reaches the
sensor decreases with the square of the distance. If you double this distance, you
receive only 1/4 of the radiation.
6-2
Humidity
30% to
10% to
Class IIa medical device according to MDD
Exposure time (msec) or pulses
Voltage (kV or kVp)
Current (mA)
Air
Pressure
700 to
95%
1060
hPa
95%
93/42/ECC
Ambient
Temperature
5° to 30°C
-40° to 70°C