KaVo IXS 1.013.9910 Manual Del Usuario página 8

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
KaVo IXS
CAUTION
Biting down on the sensor or cable will damage the housing and/or
cable and lead to a sensor malfunction.
5. Position the X-ray generator as usual for X-ray film.
Use of the paralleling technique is highly recommended.
Verify connectivity of the KaVo IXS to the computer along with
Important:
having a green driver icon before attempting to take an X-ray.
CAUTION
For maximum image quality, make sure the patient remains still
during the exposure.
WARNING
Take the necessary steps to protect yourself from radiation. For
proper operator positioning, refer to the 'Instructions for Use' of your
intraoral X-ray equipment.
6. Acquire the image by pressing the X-ray button on your radiography
system.
Important:
7. Carefully remove the sensor from the oral cavity following the exposure.
4
CORRECT
It is important that the X-ray beam hits the whole sensor
surface.
USB PORT
INCORRECT
English