Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
9.2
Processing limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
9.3
Dismantling
1. Disconnect the APC-applicator from the handpiece.
9.4
Required aids
Processing step
Items Needed
Precleaning
Soft plastic brush/soft cloth
50 ml disposable syringe with male
Luer-Lock connection
Erbe irrigation adapter 20132-087
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
10 ml disposable syringe with male
Luer-Lock connection (4 ea.)
Erbe irrigation adapter 20132-087
Compressed air gun (2 bar maximum)