OBJ_DOKU-189884-007.fm Seite 23 Freitag, 11. Juni 2021 10:59 10
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
5.2

Processing limitation

Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
5.3

Dismantling

1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.
5.4

Required aids

Processing step
Items Needed
Precleaning
Soft plastic brush/soft cloth
50 ml disposable syringe with male
Luer-Lock connection
Female-to-female luer-lock adapter